Quick Start (P001 to P007)

P001 Operation

Sets the type of measurement required for the application.

P002 Material

Specifies material type. If the material presents a flat surface perpendicular to the

transducer beam, select liquid.

P003 Maximum Process Speed

Determines level change reaction speed.

Use a setting just fast enough to keep up with your process. Slower settings provide
higher accuracy. Faster settings allow for more level fluctuations.

Values

0

Out-of-service

1

Level – how full the vessel is (Point 1 and/or 2 only)

2

Space – how empty the vessel is (Point 1 and/or 2 only)

3

*

Distance – distance from transducer to material (Point 1 and/
or 2 only)

Values

1

*

Liquid or flat surface

2

Solid or angled surface

Alters

P830 TVT Type

Values

1

Slow (0.1 m/min)

2

*

Medium (1 m/min)

3

Fast (10 m/min)

4

Surge (1.7 m/sec)

5

Immediate (17 m/sec)

Alters

P070 Failsafe Timer

P700 Max Fill Rate

P701 Max Empty Rate

P702 Filling Indicator

P703 Emptying Indicator

P704 Rate Filter

P710 Fuzz Filter

P713 Echo Lock Window

P727 Scan Delay

P841 Long Shot Number

Related

Failsafe (P070 to P072)

Rate (P700 to P708)

Measurement Verification (P710 to P713)

Transducer Scanning (P726 to P729)