SITRANS LU01 – INSTRUCTION MANUAL

Page 31

m

mmmm

Pa
ra

m

e

te

rs

P006 Empty

Enter distance in units (P005) from the face of the transducer to the process empty point.

Setting this value also sets Span (P007) unless Span was already set to another value.
For distance operation (P001=3), Span is preset to Empty.

P007 Span

Sets the range level to be measured.

Span is automatically preset to 1.1 x the blanking value (P800) less than the Empty value
(P006), unless it is altered manually.

For distance operation (P001 = 3), Span is preset to the Empty value (P006).

Enter a lower value if desired. If the automatic setting is not high enough, mount the
transducer higher (see Installation/Transducer Mounting on page 10) and enter the new
Empty (P006) distance.

Values

Range: 0.000 to 9999
Preset: 5.000 m (or equivalent depending on units)

Alters

P007 Span

Altered By

P005 Units

Related

P800 Near Blanking

P921 Material Measurement

P927 Distance Measurement

Values

Range: 0.000 to 9999
Preset: based on Empty (P006)

Alters

P112 Relay ON Setpoint

P113 Relay OFF Setpoint

Altered By

P005 Units

P006 Empty

Related

Volume (P050 to P055)

P800 Near Blanking

P921 Material Measurement

P922 Space Measurement