This feature may only be used to adjust the delay before the next point is scanned. Enter
the amount of delay in seconds. This value is automatically altered when Maximum
Process Speed (P003) is altered.

P728 Shot Delay

The delay, in seconds, between transducer shots.

Use this if transient acoustic noise within the vessel is causing measurement difficulties
due to echoes from one shot being received on the next. If more than one ultrasonic unit
is installed for redundancy, this value should be 0.

P729 Scan Time

View the elapsed time (in seconds) since the point displayed was last scanned.

This may be viewed as an Auxiliary Reading in the RUN mode.

Display (P730 to P733)

P730 Auxiliary Reading

Use this feature to display operator selected Auxiliary Readings temporarily or

indefinitely (as desired).

Select OFF to display Auxiliary Readings temporarily. Select HOLd to display Auxiliary
Readings until another Auxiliary Reading is selected or programming mode is entered.
See the

Keypad section on page 24 for RUN mode auxiliary readings.

Selecting the Auxiliary Reading operation

1.

Press READING

to display the Auxiliary Function symbol.

2.

Press ARROWS

to access the OFF or HOLd option desired.

3.

Press ENTER

.

If necessary, enter the Parameter Number to default in the Auxiliary Reading display. That
value will show in the auxiliary reading area by default. Other values are available but will
reset to the parameter defined here.

Values

Range: 0.1 to 4.0
Preset: 0.5

Values

Range: 0.000 to 9999 (view only)

Related

P001 Operation

Values

Range: 000 to 999
Display: OFF, HOLd