102

SSCI - Suburban Surgical Company, Inc.

Chapter 5 - Repairs & Replacements

5.

Un-plug the green wire from Terminal 6 on the rear of the
temperature control assembly.

6.

Un-plug the green and white wires from Terminal 4 on the
rear of the temperature control assembly.

7.

There are two terminals on the rear of the system light.
One terminal has two gray wires, the other has only one.
Dis-connect the terminal with the two gray wires.

8.

Cut both gray wires off the on/off button spade terminal.
Note: Save the spade connector, and take it to your local
hardware or electronics supply store for a replacement.

9.

Un-screw and remove the knurled ring from the rear of the
on/off button.

10.

Pull the button and attached wires out through the front of
the compartment door and discard the assembly.

Installation

1.

On the new button assembly, unscrew the knurled ring and
remove it from the assembly.

2.

From the outside of the door, pass the connectors, wires
and the button body through the mounting hole.

Figure 115. Floor Temperature On/Off Button
Replacement Part, P/N 212242

Green

White

Gray

Red

to Fuse

Holder

to System

Light

to Term 6

to Term 4

Advertising