Form No. 702718 Rev. A March, 2005

121

Owner’s Manual - Regal Intensive Care Unit

Figure 138. Oxygen Flow Meter and Related Pieces

Bracket

Support

Plate

Oxygen Tubing

Connector

Elbow Fitting

Nut

Knob

Knob Nut

7.

Before proceeding, wait 10-minutes to allow any residual
oxygen from the system to dissipate.

8.

With a Phillips screwdriver and a 3/8-in. wrench, remove
the two screws/nuts that hold the flow meter bracket to the
unit (Figure 137).

7.

Remove the flow meter and support plate.

Dis-Assembly

1.

With a 5/8-in. wrench, unscrew the elbow fitting from the
bottom of the flow meter (Figure 138).

2.

With a 1/2-in. wrench, remove the oxygen tubing connector
from the rear of the flow meter.

3.

Screw the flow meter knob counter-clockwise as far as it
will go.

4.

With a 1/2-in. wrench, unscrew the knob nut until the
knob comes free.

5.

Remove the bracket from the flow meter.