142

SSCI - Suburban Surgical Company, Inc.

Chapter 6 - Troubleshooting

Oxygen flow is excessive.

First: Make sure your oxygen pressure regulator is correctly
adjusted?

Second: Is the oxygen flow meter adjusted properly? Refer to
Controlling Oxygen Flow Rate on Page 56.

Third: Is the set point on the oxygen concentration meter
correctly set? Refer to Using Oxygen on Page 51.

Fourth: Your solenoid valve may be defective. Order P/N
853440 and refer to Page 118 for replacement instructions.

Fifth: Your oxygen flow meter may be defective. Order
P/N 853456 and refer to Page 120 for replacement instructions.

Remedial Action