Door Hinge

Left Door - P/N 212035

Right Door - P/N 212036

There is one upper and one lower hinge for each compartment door
(Figure 82). The hinges on both doors are identical and all are
removed and installed the same way. Each hinge has an upper and
lower half. Note: These instructions tell you how to replace the
lower half only. The upper half is welded to the door and cannot
be replaced other than by replacing the complete door.

Tool Required

„

Flat-blade screwdriver

Removal

1.

Open the door.

2.

Grasp the door firmly and lift up until the
upper halves of both door hinges come
free of the lower halves (Figure 83).

3.

With a flat-blade screwdriver, remove the
two mounting screws and remove the hinge
from the unit.

Installation

1.

Hold the hinge in place and loosely secure
with the two mounting screws but do not
tighten the screws down yet (Figure 83).

2.

Holding the door firmly, align the pins in the upper halves
of both door hinges with the holes in the lower halves, and
gently lower the door onto the hinges (Figure 82).

3.

Gently swing the door open and closed several times to
make sure it moves freely. This insures that the upper
and lower hinges align with each other.

4.

Tighten the hinge mounting screws.

5.

Close and latch the door.

Figure 83. Door Hinge and Mounting
Screws

Upper Half

of Hinge

Lower Half

of Hinge

Mounting

Screws

Right door, upper hinge shown; other 3 similar.