Suburban Mfg REGAL INTENSIVE CARE UNIT 12019-01-NGFREP User Manual

Page 143

Advertising
background image

Form No. 702718 Rev. A March, 2005

131

Owner’s Manual - Regal Intensive Care Unit

Sodalime Pan

P/N 206537

CAUTION: Sodalime can cause burns to eyes and
skin. Dust can cause irritation to skin and eyes on
contact. Harmful if swallowed.

CAUTION: Use safety goggles or glasses, PVC or
rubber gloves, and a nuisance dust mask when
handling Sodalime.

CAUTION: Dispose of Sodalime waste and water
rinses only in accordance with local, state and
Federal regulations.

CAUTION: For complete information, refer to the
Material Safety Data Sheet (MSDS No. 005213)
included in the Appendix to this manual.

CAUTION: Wash your hands thoroughly after
handling Sodalime.

Removal

Sodalime

Pan

Figure 155. Sodalime Pan in Filter/Fans
Compartment

1.

Open the filter/fans compartment (refer
to Page 58).

2.

Carefully, and without spilling the contents,
remove the Sodalime pan (Figure 155).

3.

Empty the Sodalime pan.

Dispose of Sodalime waste and water
rinses only in accordance with local,
state and Federal regulations.

Installation

4.

Open a single container of Sodalime.

5.

Pour the contents of the container into the new pan and
smooth it out to an even depth.

6.

Carefully, replace the filled pan into the compartment.

7.

Replace and secure the filter/fans compartment panel.

8.

When you are finished re-filling the Sodalime pan, wash
your hands thoroughly.