40

SSCI - Suburban Surgical Company, Inc.

Chapter 3 - Installation & Setup

1.

Push the main on/off switch (Figure 40) to On. The amber
light in the switch comes on.

2.

Push the cooling on/off switch (Figure 42) to On. The
amber light in the switch comes on.

3.

Turn the temperature set dial to about 5° below ambient
room temperature. Because the animal compartment is still
warm and cooling is now called for, the cooling system
activates and the red system light comes on.

4.

Observe the digital readout (Page 24) on the filter/fans
compartment front panel. The temperature slowly drops
until it reaches the level set on the temperature set dial, at
which time, the system light goes out.

5.

Push the cooling on/off switch to Off. The light in the
switch goes out.

6.

Push the main on/off switch to Off. The light in the switch
goes out.

Checking the

Cooling Function

Advertising