Chapter 3 - Installation & Setup

Connecting

Electrical Power

CAUTION: The Regal Intensive Care Unit requires a
dedicated, 15-amp minimum source of electrical power.

1.

Make sure that all seven on/off switches and buttons
(Figure 40) are in the Off position (The upper ends of all
switches should be depressed):
„

Main On/Off Switch

„

Alarm On/Off Switch

„

Oxygen On/Off Switch

„

Cooling On/Off Switch

„

Heat On/Off Switch

„

Left Floor Heat On/Off Button

„

Right Floor Heat On/Off Button

2.

Open the electronics compartment (refer to Page 57) and
remove the electrical power cord (Figure 23 on Page 20).

3.

Plug the power cord into its wall outlet.

4.

Plug the large end of the power cord into the power
connection at the rear of the unit (Figure 41) and twist
clockwise to lock.

5.

Push the main on/off switch to On. The amber light in the
switch comes on. The three circulating fans in the filter/fans
compartment and the ventilating fan in the electronics
compartment start running.

6.

Push the main on/off switch to Off. The light in the
switch goes out and all the fans stop.

Figure 40. On/Off Switches/Buttons on
Electronics Compartment Door

Main

Alarm

Oxygen

Cooling

Heat

Left Floor Heat

Right Floor Heat

Set-point

Adjustment Knob

Figure 41. Electrical Connection at Rear
of Unit

Electrical

Connection