Seal, iv port – Suburban Mfg REGAL INTENSIVE CARE UNIT 12019-01-NGFREP User Manual

Page 94

Advertising
background image

82

SSCI - Suburban Surgical Company, Inc.

Chapter 5 - Repairs & Replacements

Seal, IV Port

P/N 752100

There are four IV ports on the Intensive Care Unit; two over each
half-size compartment. Each IV port consists of a seal, a seal
retainer, two retaining screws, and two locknuts. All four IV port
seals are identical and are removed and installed the same way.

Tools Required

„

11/32-in. wrench

„

Phillips screwdriver

Removal

1.

Open both animal compartment doors.

2.

Use a 11/32-in. wrench and a Phillips screwdriver to
remove the two cap screws that hold the seal retainer in
place (Figure 91) and remove the retainer from the unit.

3.

Remove the IV Seal from the unit.

Installation

Note: On the seal and seal retainer, the mounting holes are closer
to the top edge than to the bottom edge (Figure 92).

1.

Place the IV seal behind the IV port opening with the short
side on top.

2.

Hold the seal retainer in place, flanges outward, over the
seal and secure with the two cap screws and locknuts.

3.

Close the doors.

Figure 91. Removing the IV Port

Cap Screws

Figure 92. Parts of the IV Port

Seal

Cap

Screws

Nuts

Seal
Retainer

Short

Long

Top