Setting the terminator – Konica Minolta NON-CONTACT 3D DIGITIZER VIVID 9i User Manual

Page 21

Advertising
background image

19

Connecting the VIVID 9i to a Computer

Setting the Terminator

The VIVID 9i has a built-in SCSI terminator. The terminator is required if the VIVID 9i is the last device of
those connected in series to the SCSI interface (i.e., if only one of the SCSI ports on the VIVID 9i is used).

If the terminator is not set correctly, the VIVID 9i or devices connected to the SCSI interface may mal-
function.

M

emo

“ENABLED” has been selected as the default setting before shipment.

N

ote

“DISABLE” must be selected if the VIVID 9i is connected via the SCSI interface but is not the last device of those
connected in series, and “ENABLE” must be selected if it is not connected via the SCSI interface.

[Setting Procedure]

1

Press .
The MENU view will appear.

2

Press

to locate the arrow cursor to “SC-

SI SETTINGS”.

3

Press

(or ).

The SCSI SETTINGS view will appear.

4

Press

to locate the arrow cursor to

“TERMINATE”.

5

Press

(or ).

The currently selected option (ENABLED or DISA-
BLED) will turn to pink.

6

Press

to select “ENABLED” or “DISA-

BLED”.

<Settings>
ENABLED: Select this option if the terminator is to be

used.

DISABLED: Select this option if the terminator is not to

be used.

STANDARD

3 HQ ON NF ON

COLOR OFF

W.B. DEFAULT

SCSI 5 ENABLED

WHITE BALANCE

INITIAL MODE

INITIAL CONDITION

SCSI SETTINGS

INFO

SCSI SETTINGS

SCSI ID 5

TERMINATE ENABLED

SELECT

ENTER

SCSI SETTINGS

SCSI ID 5

TERMINATE ENABLED

SELECT

ENTER

Vivid910-hard-E.book Page 19 Monday, October 18, 2004 3:11 PM