Field Calibration

Executing Field Calibration

The Field Calibration System is made up of the Field Calibration System Section and the Frame Section. It is
necessary to assemble it when it is to be used for the first time or after it has been temporarily disassembled to
move the Field Calibration System, etc. In addition, it is necessary to attach the Field Calibration System to the
main unit in order to carry out field calibration.

M

emo

VIVID 9i/VI-9i has been calibrated at 20ºC. The allowable operating environment is 10ºC to 40ºC, but we recommend
that you execute Field Calibration or scan the objects in environments where ambient temperature is 20ºC (68ºF).

Assembling the Field Calibration System

[Procedure]

1

Install the Frame Section.
Install the Frame Section in a secure location with suf-
ficient space.

N

ote

For the place that is necessary to install the Frame of the
Field Calibration System, see page 33 of this instruction
manual. A “Calibration Area” is also required for execut-
ing the calibration.

2

Adjust the Height of Rubber Feet to Stabilize the Frame Section.
The Frame of the Field Calibration System is designed to be used on a sufficiently wide, stable, flat sur-
face. However, even if a sufficiently wide space is available, the surface may not be completely flat, re-
sulting in one or more of the rubber feet of the frame not being in contact with the surface. To solve this
problem, the feet have been equipped evenly with a mechanism to allow their height to be adjusted.

M

emo

All 6 rubber feet have height adjustment mechanisms.

N

ote

• Adjust a height of the rubber feet without any item on the frame during the procedure for setting up the frame

of this system.

• When transporting this system, set the rubber feet to their lowest height to avoid damage to the shopping box.

1

Turn the lock nut counterclockwise to unlock the
rubber foot block.

Unlock the rubber
foot block by turn-
ing the lock nut
counterclockwise.