Field Calibration

6

Check that field calibration was successful.
If you are using field calibration results to measure the dimensions of the field calibration system, the
software shows the differences from the measurements made at time of shipping. Differences are shown
for four places: X1, Y1, X2, and Y2.
• If field calibration was not successful, one of the following messages appears. Proceed in accordance

with the displayed message.

7

Click the [Cancel] button.
The dialog closes, and field calibration is completed.

Message

Meaning

What To Do

Field Calibration:

Scan Error

Failed to recognize

the field calibration

system.

• Remove the laser barrier and the receiving lens cap.

• Check that you are using the correct system type for the lens

used.

• Check that foreign matter is not in the “Calibration Area”.

• Check that there is no effect of light from outside.

• Be sure that the Field Calibration Chart unit is correctly

mounted to the Frame section.

• Check that the image of the Field Calibration Chart is cor-

rectly displayed in the view angle of the finger of the VIVID
9i/VI-9i or the Work window of the dialog box of the Poly-

gon Editing Tool.

Field Calibration:
Evaluate Error

Failed to correctly
measure the field

calibration system.

• Check that you are using the correct system type for the lens

used.

• Check that foreign matter is not in the “Calibration Area”.

• Check that there is no effect of light from outside.
• Be sure that the Field Calibration Chart unit is correctly

mounted to the Frame section.

• Check that the image of the Field Calibration Chart is cor-

rectly displayed in the view angle of the finger of the VIVID

9i/VI-9i or the Work window of the dialog box of the Poly-

gon Editing Tool.

Field Calibration:

Calibrate Error

Adequate field cali-

bration results can-

not be obtained.

• Remove the lens, and then set it again correctly.

• Check that foreign matter is not in the “Calibration Area”.

• Check that there is no effect of light from outside.
• Be sure that the Field Calibration Chart unit is correctly

mounted to the Frame section.

• Check that the image of the Field Calibration Chart is cor-

rectly displayed in the view angle of the finger of the VIVID

9i/VI-9i or the Work window of the dialog box of the Poly-

gon Editing Tool.

• When multiple units of the instrument are used, check to

make sure that the Field Calibration System matches the

serial number of the instrument.

• If there is any dirt on the laser window or the receiving lens,

clean these as explained on page 3.

• Clean off any dust on the field calibration system.

• If the field calibration system is scratched or dented, contact

a KONICA MINOLTA SENSING authorized service facil-
ity.

• If the same error recurs when you retry the field calibration,

contact a KONICA MINOLTA SENSING authorized ser-
vice facility.