Replacing the Lens and Field Calibration

With the VIVID 9i, three types of lens are available: WIDE lens, MIDDLE lens and TELE lens. Be sure to
mount the lens that best matches the subject’s size and distance. Before taking scans of a subject, most suitable
lens for the size of the object and distance must be used.

N

ote

After changing the lens, you need to recalibrate using the field calibration system. For Field Calibration, refer to page
32.

Relationship between Measurement Distance and Object

Select the most suitable lens by referring to the table given below.

M

emo

The closer the object to the VIVID 9i, the higher the measurement accuracy is. On the other hand, influences of shades
produced due to roughness on the surface of the object can be reduced if the object is positioned far from the VIVID 9i.

WIDE Lens

MIDDLE Lens

TELE Lens

Unit (mm)

Measurement Distance

500

600

800

1000

2500

Object Size

Horizontal

299

359

478

598

1495

Vertical

224

269

359

449

1121

Unit (mm)

Measurement Distance

500

600

800

1000

2500

Object Size

Horizontal

165

198

263

329

823

Vertical

124

148

198

247

618

Unit (mm)

Measurement Distance

500

600

800

1000

2500

Object Size

Horizontal

93

111

148

185

463

Vertical

69

83

111

139

347