4.4.2

Calibration menu

The calibration menu is used to adjust the sensors. The following functions are available:

Menu point:

Function:

1 Cy permeate

Calibration of permeate sensor. If the menu point '1 Cy permeate' is
chosen, a submenu appears with two items: '1 min. Cy' and '2 max. Cy'.
'1 min Cy': please remove the Cy sensor and press the . key.
'2 max Cy'

Important!!! The temperature must be calibrated before calibration of the
conductivity.

Connect the Cy sensor. Now the current permeate conductivity has to be
determined by means of a standard meter, and this value has to be
entered by numerical editing. The entered value is equated with the
current conductivity after actuation of the

Í

key.

2 temp.permeate

Calibration of permeate temperature sensor.

For that purpose the current permeate temperature has to be defined
with a standard meter, and this value has to be entered by means of
numerical editing. The entered value is equated with the current
temperature when the

Í

key is then actuated.

3 calibr. reset

This menu point allows to reset adjustment of the two measuring inputs
to the internal standard values.

ATTENTION !! Internal calibration is not a correct adjustment.