Variables Table

Category

Variable Name

Description

Get

Set

Calibration_Name

Name of calibration

V

Calibration_Date

Calibration date (MMDDYY).

V

Calibration_Time

Calibration time

(HHMM, e.g., 1545).

V

Unit

Weighing unit:

ton, kg, g, lb or oz.

V

Resolution

Weighing resolution:

0.0001, 0.0002, 0.0005, 0.001,

0.002, 0.005,

0.01, 0.02, 0.05,

0.1, 0.2, 0.5,

1, 2, 5,

10, 20 or 50.

V

Full_Capacity

Full capacity of the load cell.

V

Maximum_Load

Maximum applied load on the

load cell.

V

Calibration

Info

Load_Cell_Output

Output of the load cell:

0 = Unknown

1, 2 or 3 = 1, 2 or 3 mV/V.

V

Filter1

V

*

Filter2

V

*

Filtering

Decimator

There are

two

filters. The first filter is

basically a 1st level moving average

filter of size

Filter1

(2-256). Then

depending on

Decimator

(1-1000),

every Decimator-th result from the 1st

level filter is put through another

moving average filter of size

Filter2

(2-256), which is the

second filter

.

*

S

etting

these three variables causes

a board reset, so it is a slow function.

Therefore, in order to set all of them
there is one more option: the method

Set_Filtering , which works faster

than three individual activations of

the Set_Variable method.

V

*