Omega Speaker Systems High Speed Load Cell Interface Card LCIC-WIM-BEN User Manual

Page 60

Advertising
background image

59

Notes
1. Wait 10 ms after sending the colon (‘:’) before sending the rest of the

command (‘

x

<c/r>’).

2. Wait 30 ms after sending the ‘:

x

<c/r>’ command before

analyzing the response(s).

3. There might be

four cases with the conditions of 2.1 & 2.2:

#1:

Both

conditions are true:

Both responses will be transmitted –
first ‘^

x

<c/r>’ and then ‘A

x

<c/r>’.

(Old address retired and address

x

became active.)

#2:

None of the conditions is true:

There will be no response.
(No address was active before, none is active now.)

#3: Only 2.1 is true:

‘^

x

<c/r>’ will be transmitted.

(No address is active, as old address retired and the new
one does not exist.)

#4: Only 2.2 is true:

‘A

x

<c/r>’ will be transmitted.

(Address

x

became active.)

4. Verify normal address switching by the ‘n’ command. In case of

unexpected response to the ‘n’ command, repeat the ‘:

x

<c/r>’

command. In case #1 of note 3 (which is the most frequent) this
verification is not needed, simplifying the switching procedure.

Advertising