2.2 Pneumatic Connections

1. Connect a ¼” exhaust line to the fitting labeled ‘Exhaust.’ This line should be vented
to an outside area, since the exhaust gas may still contain trace levels of ozone that may
not be completely removed by the built-in the ozone scrubber.

CAUTION

Exhaust gas from the M460H may contain dangerous levels of

ozone!


2. Connect the ozone delivery line to the ¼” inlet fitting labeled “Ozone Inlet” on the

bottom face of the enclosure (See Figure 2-2.) The ozone delivery should be
regulated to no more than 30psig. All tubing used should be made of ozone resistant
material such as PTFE(Teflon

™) or FEP. API can supply appropriate tubing for

connecting the ozone supply line.

3. Connect the oxygen or other zero gas source to the ¼” tube fitting labeled ‘Zero Gas

Inlet.’ Zero Gas supply pressure should be regulated between 5 and 30 psig.

4. The gas flow rate through the monitor should be established between 0.5 and 2.0

L/min.

Figure 2-2 Pneumatic Connections

WARNING - DO NOT EXCEED

30 PSIG GAS DELIVERY

PRESSURE

OZONE INLET

WARNING - DO NOT EXCEED

30 PSIG GAS DELIVERY

PRESSURE

ZERO GAS INLET