2.4.4 RS232/RS485 Serial Communications Port

A bi-directional RS232/485 Serial Port Interface is provided that can be used for data
acquisition and instrument control. All that is required to use this interface is a computer
with a serial communications port (COM port) and terminal software. Three pins are
provided on the I/O connector for serial communications as shown in Table 2-3.

Table 2-3 Serial Port Pins

Label

Description

SERIAL TX

Serial Transmit (RS485 – A)

SERIAL RX

Serial Receive (RS485 – B)

SERIAL GND

Serial Ground (RS232 Only)


The Serial Port can be configured for either RS232 or RS485. The standard factory
configuration is RS232. For direct connection to a PC or Laptop, RS232 should be used.
For a permanent connection into a data acquisition system, RS485 is recommended
because it has better noise immunity and is more reliable when communicating over
longer distances. See Appendix A for details on changing the Serial Port configuration.

Figure 2-5 below shows typical connections using RS232 and RS485. Note that the
RS485 connection does not use the SERIAL GND pin.

Figure 2-5 Typical RS232 and RS485 Connections


See Chapter 5 for information on the Serial Communications protocol used by the
M460H.

SERIAL TX

SERIAL RX

DB9-FEMALE

RS232

SERIAL TX

SERIAL RX

RS485-A
RS485-B

RS485

SERIAL GND