2.4.2 Digital Status Outputs

The M460H has four digital status outputs for indicating error and operational status
conditions. These outputs are in the form of opto-isolated open-collector transistors.
They can be used to drive status LED’s on a display panel or interface to a digital device
such as a Programmable Logic Controller (PLC).

Figure 2-4 shows the most common way of connecting the digital outputs to an external
device such as PLC. Note: Most devices, such as PLC’s, have internal provision for
limiting the current that the input will draw from an external device. When connecting to
a unit that does not have this feature, external dropping resistors must be used to limit the
current through the transistor output to 50mA or less.

See Chapter 7 for details on using the Status Outputs for diagnosing sensor and system-
level malfunctions.

Table 2-1 below describes the function of the status outputs.

Table 2-1 Digital Status Output Pins

Label

Name

Operation

STATUS OUT 1

Sensor O.K.

Normally On

STATUS OUT 2

Invalid Reading

Normally Off

STATUS OUT 3

Lamp Low

Normally Off

STATUS OUT 4

Cell Dirty

Normally Off

STATUS OUT 5

Auto-Zero Error

Normally Off

STATUS OUT 6

Spare

Undefined

STATUS COM

Common Pin for
all Status Outputs

N/A

Figure 2-4 Digital Output Connections

Opto-Isolator

Digital Output #1-4 (Collector)

Digital Output Common (Emmiter)

Ground Provided
by PLC

Programmable Logic
Controller or other device

M454

Digital
Input

+5V

P/N 03662D Teledyne API Model 460H O3 Monitor Operator Manual - Page 15