0 serial communications, 1 serial port configuration, 2 command syntax – Teledyne 460H User Manual

Page 21: Erial, Onfiguration, Ommand, Yntax, Table 5-1 serial port configuration

Advertising
background image

5.0 Serial Communications

The M460H comes equipped with a powerful digital Serial Communications Port that
can be used for Data Acquisition and for changing instrument configuration. This port
can be configured for either RS232 or RS485(half-duplex) operation. See Section 2.4.4
for details on connecting the port to a computer or Data Acquisition system.

5.1 Serial Port Configuration

The Serial Port of the device being used to communicate with the M460H should be
configured as follows:

Table 5-1 Serial Port Configuration

Parameter

Value

Baud Rate

9600 bps

Data Bits

8

Stop Bits

1

Parity None

5.2 Command Syntax

All characters sent and received are standard Ascii characters and all numbers are
decimal numbers converted to Ascii text.

All commands are sent using the following syntax:

<address><command>:<data1>,<data2>#<checksum><CR>

Where:
address is the instrument address(default =1)
command is the command string being sent
: (colon) is the data separator and is only included if data is being sent as part of the

command (See Command Details below to see if a command requires data or
not)

data1

is the first data parameter, if required.

data2

is the second data parameter, if required.

#

is the Checksum separator, sent only if optional checksum is included

checksum is an Ascii checksum of all characters up to the # character. The checksum is

optional, commands sent without the checksum(and checksum separator,) are
also valid.

CR

is a carriage return, Ascii 13.

P/N 03662D Teledyne API Model 460H O3 Monitor Operator Manual - Page 21

Advertising