10 vset, Syntax, Description – Teledyne 460H User Manual
Page 32: Data parameters sent, Response, Example
Advertising
5.4.10 VSET
Syntax
<address>VSET:<var_index>,<new_value><CR>
Description
Sets value of an internal configuration variable (VAR.)
Data Parameters Sent
var_index
is index number for internal VAR (See VGET for index list)
new_value
new value for VAR.
Response
<address>:<var_value>#<checksum><CR>
Example
Command:
1VSET:0,15.0<CR> - Set analog_range VAR to 15.0 (wt%)
Response:
1:OK#261<CR>
- VSET Successful
1:FAIL#391<CR>
- VSET Failed
P/N 03662D Teledyne API Model 460H O3 Monitor Operator Manual - Page 32