3 command summary, Ommand, Ummary – Teledyne 460H User Manual

Page 22: Table 5-2 serial port command summary

Advertising
background image

Examples of valid Commands with no data:
Checksum

Included:

1O3#179<CR>

No

Checksum:

1O3<CR>

Examples of valid Commands with data:
Checksum

Included:

1VSET:1,20#620<CR>
No

Checksum:

1VSET:1,20<CR>

5.3 Command Summary

Table 5-2 below lists the commands available and a summary of their function.

Table 5-2 Serial Port Command Summary

Command

Description

CAUTO

Perform an Auto-Zero Calibration (if option is installed)

CZERO

Perform a manual zero calibration (zero calibration is
computed based on current O3 reading in cell)

DACSTEP

Analog Output Test Mode, Step Function

O3

Returns current O3 concentration

SETADDR

Set communications address

TDUMP

Returns a compact list of current measurement parameters

TLIST

Returns all current measurement parameters

VGET

Get Single VAR value

VLIST

Lists all internal VARS

VSET

Sets value of internal VAR

P/N 03662D Teledyne API Model 460H O3 Monitor Operator Manual - Page 22