5.4.7 TLIST

Syntax
<address>TLIST<CR>

Description
Returns a verbose list of the current measured parameters with labels for each parameter:
O3 Concentration, Cell Pressure(psia,) Cell Temperature (K,) Lamp Temperature (K,)
Measure Detector (mV,) Calibrated Reference Detector (mV,) Reference Detector (mV)

Data Parameters Sent
None

Response
Test Parameter List (See Below)

Example
Command:

1TLIST<CR>


Response:

O3 = 0.0226168
Press = 14.7753
Cell Temp = 300.7116
Lamp Temp = 324.7965
Ref = 2881.52
Meas = 2941.092

Raw Ref = 4412.646


Notes
Note that the concentration value returned shows more digits after the decimal than the
front panel display. Concentration values should only be considered valid to 4 significant
digits.

Also, note that no checksum is sent in the response to the TLIST command.

P/N 03662D Teledyne API Model 460H O3 Monitor Operator Manual - Page 29