4.2 Auto-Zero Operation

The Auto-Zero feature allows the monitor to operate unattended for extended periods of
time. This feature automatically performs a Zero Calibration at specific intervals (default
interval is 12hrs.) The monitor does this by opening a solenoid valve to allow externally
supplied Zero Gas to purge the monitor and then automatically performs a Zero
Calibration.

4.2.1 Auto-Zero Installation

In order for this feature to function properly, external, pressurized zero gas must be
supplied to the monitor as detailed in Section 2.2.

4.2.2 Auto-Zero Function

When the Auto-Zero function is triggered, the monitor performs the steps outlined in
Table 4-1 below.

Table 4-1 Auto-Zero Function

Step

Function

Display Reads

1

Switch Auto-Zero valve on

N/A

2

Purge cell for 20 seconds

Holds and Flashes Concentration Value

3

Perform Zero Calibration

Dashes: ‘----‘

4

Switch Auto-Zero valve off

Dashes: ‘----‘

5

Hold-Off for 3 sec to return
Measurement Gas to cell

Dashes: ‘----‘

6

Return to Measurement Mode Normal Concentration Display

4.2.3 Auto Zero Error

When performing an Auto-Zero Calibration, the monitor checks to make sure the ozone
concentration is near zero before calibration. If the ozone concentration is above 0.5wt%
or 5.0 g/Nm3, the monitor will not perform the calibration and will display a “ZERO
ERR” message on the display. The monitor will also turn on the Auto-Zero Error status
signal (See Section 2.4.2 Digital Status Outputs.) This Status Signal will remain on until
another Auto-Zero Calibration is successfully performed or a Manual Zero Calibration is
performed.

If an Auto-Zero error is detected, a manual Zero Calibration should be performed (See
Section 4.1.3 Zero Calibration.)