47

CAUTION

To maintain the power supply isolation and to prevent ground loops, use an
isolated programming source when operating the power supply via remote
analog programming at the J1 connector.

6.4

REMOTE VOLTAGE PROGRAMMING OF OUTPUT VOLTAGE AND CURRENT LIMIT

Perform the following procedure to set the power supply to Remote Voltage programming:

1. Turn the power supply AC On/Off switch to Off.
2. Set setup switch SW1-1 to its UP position for output voltage external programming and

SW1-2 to its UP position for output current.

3. Set SW1, position 3 to select the programming voltage range according to Table 6-3.
4. Ensure that SW1, positions 7 and 8 are at their DOWN (default) position.
5. Connect a wire jumper between J1-8 and J1-12 (refer to Table 4-4).
6. Connect the programming source to the mating plug of J1 as shown in Fig.6-1. Observe

correct polarity for the voltage source.

7. Set the programming sources to the desired levels and turn the power supply ON. Adjust

the programming sources to change the power supply output.

NOTES:

1. SW1, positions, 4, 5, 6 and 9 are not required for remote programming. Their settings can

be determined according the application.

2. The control circuits allow the user to set the Output Voltage and Output Current up to 5%

over the model-rated maximum value. The power supply will operate within the extended
range, however it is not recommended to operate the power supply over its voltage and
current rating, and performance is not guaranteed.

SW1-3 setting

Output Voltage programming

VPGM (J1-9)

Output Current programming

IPGM (J1-10)

UP

0-10V

0-10V

DOWN

0-5V

0-5V

Table 6-3: SW1-3 setting and programming range

1

14

13

25

10

12

8

9

23

22

+

+

CURRENT LIMIT
PROGRAMMING

OUTPUT VOLTAGE

PROGRAMMING

J1 connector, rear panel view

Fig.6-1: Remote voltage programming connection