3 technical data, 1 data, Technical data – VEGA VEGASON 54V…56V User Manual
Page 16
16
VEGASON 54V … 56V
Technical data
3 Technical data
3.1 Data
Power supply
Supply voltage
is sensor supply is made from the signal condi-
tioning instrument VEGAMET, the processing
system VEGALOG or the processing system
VEGASCAN 850 (1 … 15 sensors on one two-
wire cable)
Min. sensor voltage
17 V
Power consumption
max. 1.5 W
Current consumption
90 mA
Load (signal and supply cable)
depending on the connected sensor power
per connection cable, see following load
diagrams, however max. 15
W per wire or
max. 1000 m cable length
Supply from processing system
Supply from signal conditioning instrument
VEGASCAN 850
VEGAMET or processing system VEGALOG
Measuring range
(reference plan for sensor version B … D is the transducer end. For sensor version A, the
lower flange edge is the reference plane)
VEGASON 54
1.0 … 25 m
VEGASON 55
0.8 … 45 m
VEGASON 56
- version A
1.8 … 70 m
- version B … D
1.4 … 70 m
0
5
10
15
20
25W max.
0
100
200
300
400
500
600
700
800
900
1000m
2,5 mm
2
1,5 mm
2
0,75 mm
2
1,0 mm
2
Length of the connection cable
2,5 mm
2
1,5 mm
2
0
5
10
15
20
25W max.
0
100
200
300
400
500
600
700
800
900
1000m
0,75 mm
2
1,0 mm
2
Length of the connection cable
Total power of all sensors on one
two-wire cable
Total power of all sensors on one
two-wire cable