3 measurement of liquids, Flat vessel top – VEGA VEGASON 54V…56V User Manual
Page 30
30
VEGASON 54V … 56V
< 100 mm
It is also possible to mount the sensors in
version C in a 1“ thread.
4.3 Measurement of liquids
Flat vessel top
On flat vessels, the mounting is usually done
on a very short DIN socket piece. Reference
plane is the lower edge of the flange. The
transducer should protrude out of the flange
tube.
VEGASON 55 in flange version on short
DIN socket piece
Reference plane
Mounting and installation
Min. distance
Type 54: 1 m
Type 55: 0,8 m
< 400 mm
VEGASON 56 in flange version on short
DIN socket piece
Reference plane
Min. distance
1,8 m
A mounting location directly on the vessel top
is ideal. A round opening in the vessel top is
sufficient to fasten the sensor with the flange,
or version B and C with swivelling holder.
Flange version and swivelling holder on flat vessel
top
Reference
plane
Reference
plane
< 60 mm
Mounting of the transducer in 1“ thread
Reference plane
Min. meas. distance
1,8 m
Min. meas. distance
1,4 m
Swivelling
holder