Strong heat fluctuations, Min. distance to the medium, Sensor too close to the vessel wall – VEGA VEGASON 54V…56V User Manual

Page 37

Advertising
background image

VEGASON 54V … 56V

37

2 m

4 m

6 m

8 m

10 m

20 m

30 m

B

A

max. meas.
distance

Distance of the
transducer to the
vessel wall

Curve 1 (liquids)

Curve 2 (solids)

If the sensor is mounted too close to the
vessel wall (dimension A in diagram), strong
false echoes can be caused. Buildup, rivets,
screws or weld joints on the vessel wall su-
perimpose their echoes to the product or
useful echo. Please ensure the sufficient
distance of the sensor to the vessel wall,
depending on the maximum measuring dis-
tance (dimension B in diagram). In case of
good reflection conditions (liquids, no vessel
installations), we recommend determining the
sensor distance according to

Diagram curve

1

. At a max. meas. distance of e.g. 10 m, the

distance of the transducer (according to
curve 1) should be approx. 1.5 m. In case of
solids with bad reflection properties, deter-
mine the distance to the vessel wall accord-
ing to

Diagram curve 2

. Under very bad

measuring conditions (rough vessel walls,
struts), it might be necessary, to increase the
distance to the vessel wall, or to filter out the
false echoes in addition by storing them in
memory, thereby adapting the sensor more
precisely to the environment.

Mounting and installation

Strong heat fluctuations

Strong heat fluctuations, e.g. due to the sun,
cause measuring errors. Please provide a
sun shield.

Min. distance to the medium

If the min. distance to the medium is not main-
tained, the instruments show wrong meas-
ured values. Mount the instrument at the
required min. distance.

Sensor too close to the vessel wall

Sensor too close to the vessel wall

Correct

Wrong

Strong heat fluctuations

Shield

Advertising