Adjustment steps, Configuration of the input, Configuration of the measurement loop – VEGA VEGASON 54V…56V User Manual
Page 68: Measurement in gases, Adjustment structure, Setup - adjustment with vegamet or minicom, See menu schematic page 74), Display of meas. values main menu
68
VEGASON 54V … 56V
Setup - Adjustment with VEGAMET or MINICOM
Adjustment steps
On page 72 … 79 you can find the complete
menu schematic of the VEGAMET 515V si-
gnal conditioning instrument as well as of the
adjustment module MINICOM. The menu
schematic of the VEGAMET 514V signal
conditioning instrument is nearly identical. Set
up the sensor in the numbered sequence.
The respective numbers are stated in the
menu schematics. The items in brackets can
be adjusted beside the VEGAMET signal
conditioning instrument also with the adjust-
ment module MINICOM.
1.
Configuration of the input (as a rule, al-
ready preconfigured)
2.
Configuration of the measurement loop
(as a rule, already preconfigured)
3.
Measurement in gases
4.
Operating range
5.
Meas. conditions
6.
Min/Max adjustment
7.
Conditioning/Scaling VEGAMET
8.
False echo memory (only required when
errors occur during operation)
9.
Indication of the useful and noise level
10. Conditioning/Scaling of the sensor dis-
play
11. Outputs
Short explanations to the setup steps 1 … 11
follow.
1. Configuration of the input
First of all, you have to inform the signal con-
ditioning instrument (only VEGAMET 515V) to
which input (input 1 or 2) the sensor is con-
nected.
2. Configuration of the measurement
loop
(see menu schematic page 74)
3. Measurement in gases
Adjustment is only necessary when the
measurement is made in gases (Co
2
, He,
etc.) deviating from air. In case of measure-
ment in gases, sound the distance of the
sensor to the product surface and enter it in
the menu item "Measurement in gases“. The
sensor can then take the modified sonic
velocity in gases (as opposed to air) and
output correct levels.
Adjustment structure
ESC
OK
OK
OK
OK
ESC
ESC
TAG1
%
36.9
TAG2
%
TAG3
%
Menu window
Parameter
Value
weite-
re
Funk-
tion
Konfi-
gura-
tion
Param.
TAG3
Param.
TAG2
Param.
TAG1
Passwort
a u s
Ausgänge
Meß-
s t e l l e
Eingänge
Simu-
l a t i o n
Ausgang
Auswertung
Abgleich
Skalie-
rung
Lin.
K u r v e
Integra-
tions-
z e i t
10
Display of
meas.
values
Main
menu
Selection of the menu window int he
horizontal menu level as well as
selection of fixed parameters
MST1-
2