3 adjustment – VEGA VEGASON 54V…56V User Manual

Page 6

Advertising
background image

6

VEGASON 54V … 56V

2

2

Product description

1.3 Adjustment

Each measuring situation is unique. For that
reason, every ultrasonic sensor needs some
basic information on the application and the
environment, e.g. an empty vessel profile is
important for a reliable measurement. Beside
this, many other settings and adjustments
are possible on VEGASON ultrasonic sen-
sors.

The adjustment and parameter setting of
ultrasonic sensors are carried out with
- the PC
- the detachable adjustment module

MINICOM

- the VEGAMET 514V or VEGAMET 515V

signal conditioning instrument

Adjustment with PC

The setup and adjustment of the ultrasonic
sensors is generally done on the PC with the
adjustment program VEGA Visual Operating
(VVO) under Windows

®

.

The program leads

quickly through the adjustment and param-
eter setting by means of pictures, graphics
and process visualisations.

The PC can be connected to any individual
position of the system or the signal cable. It is
connected by means of the two-wire PC
interface converter VEGACONNECT 2 to the
sensor, the signal cable or to the signal con-
ditioning instrument or directly with a stand-
ard cable (RS 232) to the VEGALOG
processing system.

The adjustment and parameter data can be
saved with the adjustment software on the
PC and can be protected by passwords. On
request, the adjustments can be quickly
transferred to other sensors.

Adjustment with the PC to the digital signal and
supply cable between the sensors and the VEGAMET
signal conditioning instrument or on the sensor itself