13 special function: manual correction, Settings in the "extended menu – VEGA VEGAMET 514V User Manual

Page 42

Advertising
background image

42

VEGAMET 514V

1)

Settings in the "extended menu“

7.13

Special function: Manual correction

Manual offset correction

(only with hydrostatic pressure transmitters)
This correction can be carried out to com-
pensate modifications of the sensor values
due to ageing. In order to do this, the sensor
must be in an unpressurized condition
(empty vessel).

Param.
TAG-
No. 2

like TAG-No. 1

Param.
TAG-
Nr. 1

TAG -
Nr. 1
%
xx,x

Adjust-
ment

Signal
condi-
tioning

Out-
puts

Simu-
lation

Spe-
cial
func-
tions

Manu-
al
corr.

Offset
cor-
rect-
ion

Offset
correc-
tion
OK?

Correc-
tion
now!
OK ?

Correc-
tion
now!
OK ?

Real
value
corr.

Real
value
corr.
OK?

Correc-
tion
%

0,0

Manual real value correction

In the menu item "Correction % 0.0“ you can
enter a percentage value which corresponds
to the actual level.

Reset
Level

Reset

OK ?

Reset
now!

OK ?

Simu-
lation
%

XX,X

Simu-
lation
now!
OK?

Real
value
corr.
at %

0

Density
corr.
at %

0

1)

1)

This menu item only appears if with the measure-
ment loop configuration a correction mode was
adjusted. Then here the percentage value must be
adjusted at which the automatic correction is
carried out.