13 special function: manual correction, Settings in the "extended menu – VEGA VEGAMET 514V User Manual
Page 42
42
VEGAMET 514V
1)
Settings in the "extended menu“
7.13
Special function: Manual correction
Manual offset correction
(only with hydrostatic pressure transmitters)
This correction can be carried out to com-
pensate modifications of the sensor values
due to ageing. In order to do this, the sensor
must be in an unpressurized condition
(empty vessel).
Param.
TAG-
No. 2
like TAG-No. 1
Param.
TAG-
Nr. 1
TAG -
Nr. 1
%
xx,x
Adjust-
ment
Signal
condi-
tioning
Out-
puts
Simu-
lation
Spe-
cial
func-
tions
Manu-
al
corr.
Offset
cor-
rect-
ion
Offset
correc-
tion
OK?
Correc-
tion
now!
OK ?
Correc-
tion
now!
OK ?
Real
value
corr.
Real
value
corr.
OK?
Correc-
tion
%
0,0
Manual real value correction
In the menu item "Correction % 0.0“ you can
enter a percentage value which corresponds
to the actual level.
Reset
Level
Reset
OK ?
Reset
now!
OK ?
Simu-
lation
%
XX,X
Simu-
lation
now!
OK?
Real
value
corr.
at %
0
Density
corr.
at %
0
1)
1)
This menu item only appears if with the measure-
ment loop configuration a correction mode was
adjusted. Then here the percentage value must be
adjusted at which the automatic correction is
carried out.