VEGA VEGASON 71 - 2 … 75 - 2 User Manual

Page 20

Advertising
background image

Adjustment at level measurement

20

71 - 2 … 75 - 2

Level measurement

Two separate level measurements can be obtained
with the VEGASON 71 - 2 … 75 - 2 (measuring
system).

The adjustment is carried out for
-

level measurement 1 = measuring channel 1
in mode range no. 10 … 12

-

level measurement 2 = measuring channel 2
in mode range no. 20 … 22

The adjustment procedures are carried out sepa-
rately for each sensor.

The display coordination is programmable and
supplied as factory preset in mode 40 as follows:

0 – 4 0

1 1

level 2 (display 2)
level 1 (display 1)

Selection of adjustment procedure

The type of empty adjustment (mode 11) and full
adjustment (mode 12) is determined with the para-
meter adjustment of this mode.

The type of adjustment depends on the local
conditions.

Mode description

Mode 10 = 1

means adjustment in m without product level change, i.e.
- with this adjustment procedure the distances in m determined from the vessel drawings

are programmed relating to the levels 0 % and 100 %

- level is not important

Mode 10 = 3

means adjustment in % without product level change, i.e.
- with this adjustment procedure the levels are programmed in % and the respective

measured distance (sensor … level) is calculated automatically

- product level change is needed.

Programming example

Change of mode 10 = 3 of mode 10 = 1 ..................

• select mode 0 - 10 in MODEFIELD
• program figure1 in PARAMETER FIELD
• then store with STO-button

Same procedure for channel 2, however mode 0 - 20.

Demonstration

Mode 10 = 1

Mode 10 = 3

Min. distance
^ 100 %

partly filling
= e.g. 75 %

partly emptying
= e.g. 25 %

Max. distance
^ 0 %

MODE 0 - 10

Advertising