Measurement, Current output

71 - 2 … 75 - 2

37

The pulse-echo measuring system can be equipped
with a current module.

The module has two current outputs.

Coordination

MODE 0 - 64

Each current output can be coordinated separately
to either channel (or difference).

Same procedure for current output 2, but with
respective modes, see above.

Mode coordination / meaning

current

coordination

range of characteristics

output

final value

initial value

1

0 - 64 = 1

channel 1

2

channel 2

0 - 65

0 - 66

3

difference

2

0 - 68 = 1

channel 1

2

channel 2

0 - 69

0 - 70

3

difference

Programming example
Coordinate current output 1, channel 2 ....................

• select mode 0 - 64 in MODEFIELD
• program figure 2 in PARAMETER FIELD
• then store with STO-button

Output characteristic

MODE 0 - 65 and 0 - 66

The characteristic of the current output is fixed by
an initial and final point.

The factory preset defines a course of 4 … 20 mA
acc. to 0 … 100 %.

Within the total range of 0.00 … 20.00 mA each
current value can be programmed for a rising or
falling level.

The current-

between initial and final value must

be min. 1 mA.

• select mode 0 - 65 in MODEFIELD
• program 16.00 mA in PARAMETER FIELD
• then store with STO-button

• select mode 0 - 66 in MODEFIELD
• program 00.00 mA in PARAMETER FIELD
• then store with STO-button

%
100

0

mA
20

4

End of characteristic
100 % ^ 16 mA

Start of characteristic
0 % ^ 0 mA

Factory preset

Programming example

Module for current output 1 and 2