VEGA VEGASON 71 - 2 … 75 - 2 User Manual

Page 32

Advertising
background image

Measurement, Relay output

32

71 - 2 … 75 - 2

Mode

(Module 1, relay output 1) continuation

Mode 0 - 44 = 1
two-point level switch
switching command 1 via
switching command 2
correspond to function B
(protection against dry
running of pumps)

Switching command 1 and 2

(Module 1, relay output 1)

MODE 0 - 45 and 0 - 46

Mode 0 - 43 defines the measurement units
coordinated to the relay output.

Mode 0-44 determines the function of the relay
output.

Mode 0-45 = switching command 1 corresponds

relay switched on (on)

Mode 0-46 = switching command 2 corresponds

relay switched off (off)

With the operating sequence of the switching
commands, the A- or B-function i.e. overfill protec-
tion or protection against dry running of pumps is
determined.
See page 31 and 32.

The min.

for the switching commands is 10 cm or

0,5 %.

Programming example for relay output 1 ...............

-

coordination to channel 1 with units in %

-

mode 1

-

two-point level switch in function A
(see page 31)

Coordination (channel 1 in %) ..................................

• select mode 0 - 43 in MODEFIELD
• program figure 2 in PARAMETER FIELD
• then store with STO-button

Mode (two-point level switch) ....................................

• select mode 0 - 44 in MODEFIELD
• program figure 1 in PARAMETER FIELD
• then store with STO-button

Switching command 1 (on) ......................................

• select mode 0 - 45 in MODEFIELD
• program 025.0 % in PARAMETER FIELD
• then store with STO-button

Switching command 2 (off) .......................................

• select mode 0 - 46 in MODEFIELD
• program 050.0 % in PARAMETER FIELD
• then store with STO-button

Same function and procedure
for module 1, relay output 2
and module 2, relay output 3 and 4, however
with respective modes, see page 31.

0

25

75

100

50

% m

distance acc. to sensor type

switching command 1,
Mode 0-45

switching command 2,

Mode 0-46

(off)

(on)