Code
Text mes-
sage

Cause

Rectification

F125

Unpermissible
electronics
temperature

l Temperature of the elec-

tronics in the non-specified
section

l Check ambient tempera-

ture

l Isolate electronics
l Use instrument with higher

temperature range

F260

Error in the
calibration

l Error in the calibration car-

ried out in the factory

l Error in the EEPROM

l Exchange of the elec-

tronics

l Send instrument for repair

F261

Error in the
configuration

l Error during setup
l False signal suppression

faulty

l Error when carrying out a

reset

l Repeat setup
l Repeat reset

F264

Installation/S-
etup error

l Adjustment not within the

vessel height/measuring
range

l Max. measuring range of

the instrument not suffi-
cient

l Check or correct installa-

tion and/or parameter ad-
justment

l Use an instrument with

bigger measuring range

F265

Measurement
function dis-
turbed

l Sensor does no longer

carry out a measurement

l Operating voltage too low

l Check operating voltage
l Carry out a reset
l Separate operating voltage

briefly

The following table shows the error codes and text messages in the
status message "Function check" and provides information on causes
as well as corrective measures.

Code
Text mes-
sage

Cause

Rectification

C700

Simulation

l Simulation active

l Finish simulation
l Wait for the automatic end

after 60 mins.

The following table shows the error codes and text messages in the
status message "Out of specification" and provides information on
causes as well as corrective measures.

Function check

Out of specification

LEVEL TRANSMITTER 8136 • 4 … 20 mA/HART - two-wire

51

9 Diagnosis, Asset Management and service