Burkert Type 8136 User Manual

Page 62

Advertising
background image

0,5 m (1.6 ft)

10 mm (0.394 in)

- 10 mm (- 0.394 in)

2 mm (0.079 in)

- 2 mm (- 0.079 in)

0

1 2

3

Fig. 42: Deviation under reference conditions - encapsulated antenna system

1

Reference plane

2

Antenna edge

3

Recommended measuring range

0,5 m (1.6 ft)

10 mm (0.394 in)

- 10 mm (- 0.394 in)

2 mm (0.079 in)

- 2 mm (- 0.079 in)

0

1

2

3

Fig. 43: Deviation under reference conditions - plastic horn antenna

1

Reference plane

2

Antenna edge

3

Recommended measuring range

Reproducibility

≤ ±1

mm

Deviation with bulk solids

The values depend to a great extent on the
application. Binding specifications are thus not
possible.

Variables influencing measurement accuracy
Specifications apply to the HART signal and the current output

Temperature drift - Digital output

±3

mm/10 K relating to the max. measuring range

or max. 10 mm

62

LEVEL TRANSMITTER 8136 • 4 … 20 mA/HART - two-wire

11 Supplement

41783

-EN

-120316

Advertising