Burkert Type 8136 User Manual

Page 52

Advertising
background image

Code
Text mes-
sage

Cause

Rectification

S600

Unpermissible
electronics
temperature

l Temperature of the elec-

tronics in the non-specified
section

l Check ambient tempera-

ture

l Isolate electronics
l Use instrument with higher

temperature range

S601

Overfilling

l Danger of vessel overfilling l Make sure that there is no

further filling

l Check level in the vessel

The following table shows the error codes and text messages in the
status message "Maintenance" and provides information on causes as
well as corrective measures.

Code
Text mes-
sage

Cause

Rectification

M500

Error with the
reset delivery
status

l With the reset to delivery

status, the data could not
be restored

l Repeat reset
l Load XML file with sensor

data into the sensor

M501

Error in the
non-active lin-
earization ta-
ble

l Hardware error EEPROM

l Exchange of the elec-

tronics

l Send instrument for repair

M502

Error in the
diagnosis
memory

l Hardware error EEPROM

l Exchange of the elec-

tronics

l Send instrument for repair

M503

Reliability too
low

l The echot/noise ratio is the

small for a reliable meas-
urement

l Check installation and

process conditions

l Clean the antenna
l Change polarisation direc-

tion

l Use instrument with higher

sensitivity

M504

Error on an
device inter-
face

l Hardware defect

l Check connections
l Exchange of the elec-

tronics

l Send instrument for repair

Maintenance

52

LEVEL TRANSMITTER 8136 • 4 … 20 mA/HART - two-wire

9 Diagnosis, Asset Management and service

41783

-EN

-120316

Advertising