Burkert Type 8136 User Manual

Page 64

Advertising
background image

Vessel pressure

-

Encapsulated antenna system

-1 … 3 bar (-100 … 300 kPa/ -14.5 … 43.5 psi)

-

Plastic horn antenna

-1 … 2 bar (-100 … 200 kPa/-14.5 … 29.0 psig)

-

Version with adapter flange from DN 150
PP

-1 … 1 bar (-100 … 100 kPa/-14.5 … 14.5 psig)

Process temperature (measured on the
process fitting)

-40 … +80 °C (-40 … +176 °F)

Vibration resistance

-

With adapter flange

2 g at 5 … 200 Hz according to EN60068-2-6
(vibration with resonance)

-

with mounting strap

1 g at 5 … 200 Hz according to EN60068-2-6
(vibration with resonance)

Shock resistance

100 g, 6 ms according to EN60068-2-27 (mechan-
ical shock)

Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar

Cable entry/plug

5)

-

Single chamber housing

l

1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm),
1 x blind stopper M20 x 1.5

or:

l

1 x closing cap M20 x 1.5; 1 x blind stopper
M20 x 1.5

or:

l

1 x closing cap ½ NPT, 1 x blind plug ½ NPT

or:

l

1 x plug (depending on the version), 1 x blind
stopper M20 x 1.5

-

Double chamber housing

l

1 x cable entry M20 x 1.5 (cable: ø 5 … 9 mm),
1 x blind stopper M20 x 1.5; 1 x blind stopper
M16 x 1.5 or optional

6)

1 x plug M12 x 1 for

external indicating and adjustment unit

or:

l

1 x closing cap ½ NPT, 1 x blind stopper
½

NPT, 1 x blind stopper M16 x 1.5 or

optionally

7)

1 x plug M12 x 1 for external

indicating and adjustment unit

or:

l

1 x plug (depending on the version), 1 x blind
stopper M20 x 1.5; 1 x blind stopper M16 x 1.5
or optionally

8)

1 x plug M12 x 1 for external

indicating and adjustment unit

5)

Depending on the version M12 x 1, according to ISO 4400, Harting, 7/8" FF.

6)

Only with Ex d version

7)

Only with Ex d version

8)

Only with Ex d version

64

LEVEL TRANSMITTER 8136 • 4 … 20 mA/HART - two-wire

11 Supplement

41783

-EN

-120316