Comm checking buttons – SKC DataTrac for the Leland Legacy User Manual

Page 14

Advertising
background image

11

SKC Real Time Monitor

Set the pump to 10 L/min. If the calibrator displays 9.7 L/min, click the  butt on
in Flow Calibrate until the calibrator displays 10 L/min. If the calibrator displays
10.5 L/min, click the  button in Flow Calibrate until the calibrator displays
10 L/min. Repeat calibration aft er sampling to verify fl ow.

Comm Checking Buttons

The Comm Checking butt ons (Figure 9) turn the communication checking function
on or off . Comm Checking monitors the interface cable connection between the PC
and the pump. The default value is On. If the interface cable becomes detached, an
error message displays (Figure 10). Reconnect the pump and click on Retry. If the
Off butt on is selected in Comm Checking, the pump’s real time information will
not be updated.

If programming more than one pump, turn Comm Checking off by clicking the
Off butt on. Turning Comm Checking off when programming multiple pumps will
eliminate the error message that displays each time the pump is disconnected.

When Comm Checking is turned off , the pump will enter Sleep mode fi ve minutes
aft er the last interaction between the computer and the pump.

Figure 9.

Comm Checking Button

Figure 10.

Error Message

Advertising