Test Port Output Characteristics

Frequency characteristics
Range

300 kHz to 3 GHz

Option 006

30 kHz to 6 GHz

Resolution

1 Hz

Stability

typically ±7.5 ppm 0° to 55° C
typically ±3 ppm/year

Option 1D5

typically ±0.05 ppm 0° to 55° C
typically ±0.5 ppm/year

Accuracy

±10 ppm at 25° ± 5° C

Power range

-5 to +20 dBm

Option 006

-5 to +18 dBm

Resolution

0.01 dB

Level accuracy

1,2

±1.0 dB

Level linearity

1,2,5

±0.25 dB, -5 to +15 dBm
±0.5 dB, +15 to +20 dBm

Impedance

50

nominal

300 kHz to 3 GHz

2

>16 dB RL (<1.38 SWR)

300 kHz to 6 GHz

>14 dB RL (<1.50 SWR)

Spectral purity
2nd harmonic

3

<-40 dBc at +10 dBm

6

<-50 dBc at 0 dBm

6

Option 002

<-25 dBc at max power

3rd harmonic

4

<-40 dBc at +10 dBm

6

<-50 dBc at 0 dBm

6

Option 002

<-25 dBc at max power

Nonharmonic spurious

6

Mixer related

<-30 dBc at max power
<-55 dBc at -10 dBm

Test Port Input Characteristics Option 011

Frequency range

300 kHz to 3 GHz

Option 006

30 kHz to 6 GHz

Average noise level

2

50 kHz to 3 GHz

<-90 dBm (3 kHz BW)

50 kHz to 3 GHz

<-110 dBm (10 Hz BW)

50 kHz to 3 GHz

<-120 dBm (10 Hz BW) typical

3 to 6 GHz

<-85 dBm (3 kHz BW)

3 to 6 GHz

<-105 dBm (10 Hz BW)

3 to 6 GHz

<-115 dBm (10 Hz BW) typical

Maximum input level

0 dBm

Damage level

20 dBm or 25 VDC

Impedance

50 ohms nominal

300 kHz to 2 MHz

20 dB RL

2 MHz to 1.3 GHz

24 dB RL

1.3 GHz to 3 GHz

19 dB RL

Option 006

3 GHz to 6 GHz

15 dB RL

6

Frequency response
300 kHz to 3 GHz

±1.0 dB

3 GHz to 6 GHz

±2.0 dB

Harmonics (Option 002)
2nd harmonic

3

<-15 dBc at 0 dBm
<-30 dBc at -10 dBm

6

<-45 dBc at -30 dBm

6

3rd harmonic

4

<-30 dBc at 0 dBm
<-50 dBc at -10 dBm

6

<-50 dBc at -30 dBm

6

Harmonic measurement accuracy
16 MHz to 3 GHz

±1.5 dB

3 GHz to 6 GHz Option 006

±3 dB

Harmonic measurement dynamic range
-40 dBc

6

(output = -10 dBm, input <-15 dBm)

Agilent 8753ES Option 011 Specifications

1. +10 dBm output power for Agilent 8753ES Option 011.
2. Typical below 300 kHz.
3. 16 MHz to 3 GHz.
4. 16 MHz to 2 GHz.
5. For Agilent 8753ES Option 011 and Option 006, linearity is specified for the

ranges of -5 to +13 dBm and +13 to +18 dBm.

6. Typical performance.