Specifications – Agilent Technologies 8753ET User Manual

Page 23

Advertising
background image

23

Agilent 85047A S-Parameter Test Set

The 85047A S-parameter test set provides the
capability to simultaneously measure the reflec-
tion and transmission characteristics of two-port
devices in either direction with a single connec-
tion. This test set includes a frequency doubler
that can be switched in by an 8753B/C Option 006
to measure 3 MHz to 6 GHz in a single sweep or
switched out to measure 300 kHz to 3 GHz in
a single sweep. The 8753ES Option 011 does not
use the frequency doubler, so the full 300 kHz to
6 GHz range is available. This test set exhibits
<5 dB insertion loss between the RF input and the
test ports for as high as 15 dBm at the test port,
and also includes a programmable step attenuator.
There are two rear panel BNC outputs. One pro-
vides a TTL signal which indicates the result of
a limit test. The second TTL output is controlled
from the 8753ES test sequence function.

Specifications

Impedance

50 ohms

Frequency range

300 kHz to 3 GHz and
3 GHz to 6 GHz
(8753B/C);
300 kHz to 6 GHz
(8753D/E/ES Opt. 006)

Directivity

1

300 kHz to 1.3 GHz

35 dB

2

1.3 GHz to 3 GHz

30 dB

3 GHz to 6 GHz

25 dB

Typical tracking

1

Transmission magnitude, phase

3

300 kHz to 3 GHz

±1.5 dB, ±10°

3 GHz to 6 GHz

+0.5, -2.5 dB, ±20°

Reflection magnitude, phase

3

300 kHz to 3 GHz

±1.5 dB, ±10°

3 GHz to 6 GHz

±1.5 dB, ±20°

Source match

1

300 kHz to 1.3 GHz

20 dB

1.3 GHz to 3 GHz

16 dB

3 GHz to 6 GHz

14 dB

Normal insertion loss
Input to port 1,2

300 kHz to 3 GHz

4.0 dB +0.8 dB/GHz

3 GHz to 6 GHz

17.5 dB +0.8 dB/GHz

Input to R

300 kHz to 3 GHz

19 dB +0.5 dB/GHz

3 GHz to 6 GHz

34 dB +0.5 dB/GHz

Port 1,2 to A,B

16 dB

Typical isolation

4

300 kHz to 3 GHz

100 dB

3 GHz to 6 GHz

90 dB

Test port switch

±0.03 dB; ±0.01 dB4

repeatability

Maximum operating level

+20 dBm

Damage level

+30 dBm

RF attenuator range

70 dB (10 dB steps)

DC bias range

±30 VDC, 200 mA,
no degradation in RF specs,
1A max.

RF connectors
Port 1,2

7 mm precision

All others

50 ohm type-N(f)

Dimensions

90 mm H x 432 mm W x
553 mm D

Weight

10 kg (22 lb)

A standard 85047A test set contains a solid-state
transfer switch, which allows continuous switching
of power from port 1 to port 2 for full two-port
error correction.

1. This can be greatly improved with accuracy enhancement.
2. Some degradation at environmental extremes below 600 kHz.
3. Degrees, specified as deviation from linear phase.
4. Typical performance.

Agilent 85047A Schematic

Advertising
This manual is related to the following products: