Measurement port characteristics – Agilent Technologies 8753ET User Manual

Page 7

Advertising
background image

Agilent 8753ES (75-Ohm Systems)

Type-N test ports

Configuration
Network analyzer

8753ES Option 075

Calibration kit

85036B

Test-port cables

11857B

System Dynamic Range
30 kHz to 50 kHz

68 dB

1

50 kHz to 300 kHz

90 dB

1

300 kHz to 16 MHz

96 dB

16 MHz to 3 GHz

106 dB

Measurement Port Characteristics

Frequency Range

Corrected

30 kHz–300 kHz

2

300 kHz–1.3 GHz 1.3 GHz–3 GHz

Directivity

48 dB

48 dB

43 dB

Source match

47 dB

41 dB

35 dB

Load match

48 dB

48 dB

43 dB

Reflection

±(0.004 dB

±(0.010 dB

±(0.019 dB

tracking

+0.02 dB/°C)

+0.01 dB/°C)

+0.02 dB/°C)

Transmission

±(0.018 dB

±(0.016 dB

±(0.033 dB

tracking

+0.02 dB/°C)

+0.01 dB/°C)

+0.02 dB/°C)

Uncorrected

2

Directivity

20 dB*

35 dB

30 dB

Source match

16 dB**

16 dB

16 dB

Load match

15 dB**

18 dB

16 dB

Reflection

±2.5 dB

±1 dB

±1 dB

tracking

Transmission

±2.5 dB

±1 dB

±1 dB

tracking

Crosstalk

90 dB***

100 dB

100 dB

*15 dB, 30 to 50 kHz

**10 dB, 30 to 50 kHz

***60 dB, 30 to 50 kHz

7

1. Typical below 300 kHz.
2. Typical performance.

Transmission uncertainty

Magnitude

Phase

Magnitude

Phase

Reflection uncertainty

Advertising
This manual is related to the following products: