System performance summaries, Measurement port characteristics – Agilent Technologies 8753ET User Manual

Page 4

Advertising
background image

4

Agilent 8753ES (50-ohm systems)

7-mm test ports

Configuration
Network analyzer

8753ES
Standard and Option 006

Calibration kit

85031B

Test-port cables

11857D

System dynamic range
30 kHz to 50 kHz

70 dB

1

50 kHz to 300 kHz

90 dB

1

300 kHz to 16 MHz

100 dB

16 MHz to 3 GHz

110 dB

Option 014

108 dB

3 GHz to 6 GHz

105 dB

Option 014

103 dB

Measurement Port Characteristics

Frequency Range

Corrected

30 kHz–

300 kHz–

1.3 GHz–

3 GHz–

300 kHz

2

1.3 GHz

3 GHz

6 GHz

Directivity

55 dB

55 dB

51 dB

46 dB

Source match 55 dB

51 dB

49 dB

43 dB

Load match

55 dB

55 dB

51 dB

46 dB

Reflection

±(0.001 dB

±(0.001 dB

±(0.005 dB

±(0.020 dB

tracking

+0.02 dB/°C) +0.01 dB/°C) +0.02 dB/°C) +0.03 dB/°C)

Transmission ±(0.008 dB

±(0.006 dB

±(0.009 dB

±(0.021 dB

tracking

+0.02 dB/°C) +0.01 dB/°C) +0.02 dB/°C) +0.03 dB/°C)

Uncorrected

Directivity

20 dB*

35 dB

30 dB

25 dB

Source match 18 dB**

16 dB

16 dB

14 dB

Load match

18 dB**

18 dB

16 dB

14 dB

Reflection

±2.5 dB

±1 dB

±1 dB

±1.5 dB

tracking

Transmission ±2.5 dB

±1 dB

±1 dB

±1.5 dB

tracking

Crosstalk

90 dB***

100 dB

100 dB

90 dB

*15 dB, 30 kHz to 50 kHz

**10 dB, 30 kHz to 50 kHz

***60 dB, 30 kHz to 50 kHz

System Performance Summaries

1. Typical below 300 kHz.
2. Typical performance.

Reflection uncertainty

Magnitude

Phase

Magnitude

Phase

Transmission uncertainty

Advertising
This manual is related to the following products: