Test-Port Output Characteristics

Frequency characteristics
Range

300 kHz to 3 GHz

Option 006

300 kHz to 6 GHz

Resolution

1 Hz

Stability

4

±7.5 ppm (0° to 55° C)
±3 ppm/year

Option 1D5

±0.05 ppm (0° to 55° C)
±0.5 ppm/year

Accuracy

±10 ppm at 25° ±5° C

Power range

-20 to +5 dBm
-85 to +10 dBm
(with Option 004)

Resolution

0.01 dB

Level accuracy

1

±1.0 dB

Level linearity

1

Standard

-20 to -15 dBm

±0.5 dB

-15 to 0 dBm

±0.2 dB

0 to +5 dBm

±0.5 dB

Option 004

-15 to +5 dBm

±0.2 dB

+5 to +10 dBm

±0.5 dB

Impedance

50

(nominal)

300 kHz to 3 GHz

>18 dB RL (< 1.28 SWR)

3 GHz to 6 GHz

>14 dB RL (<1.50 SWR)

Spectral purity
2nd harmonic

2

<-40 dBc at 0 dBm

4

<-50 dBc at -10 dBm

4

Option 002

<-25 dBc at maximum output
power

3rd harmonic

3

<-40 dBc at 0 dBm
<-50 dBc at -10 dBm

Option 002

<-25 dBc at maximum output
power

Nonharmonic spurious
Mixer related

<-30 dBc at +10 dBm
<-55 dBc at -10 dBm

Test Port Input Characteristics

Frequency range

300 kHz to 3 GHz

Option 006

300 kHz to 6 GHz

Average noise level
300 kHz to 3 GHz

<-90 dBm (3 kHz BW)

300 kHz to 3 GHz

<-110 dBm (10 Hz BW)

300 kHz to 3 GHz

<-120 dBm (10 Hz BW)

typical

3 GHz to 6 GHz

<-85 dBm (3 kHz BW)

3 GHz to 6 GHz

<-105 dBm (10 Hz BW)

3 GHz to 6 GHz

<-114 dBm (10 Hz BW)

typical

Maximum input level

+10 dBm reflection port
0 dBm transmission port

Damage level

>+26 dBm or 35 VDC

Impedance

50

(nominal)

Frequency response

5

300 kHz to 3 GHz

±1.0 dB

3 GHz to 6 GHz

±2.0 dB

Harmonics (Option 002)
2nd harmonic

2

<-15 dBc at +8 dBm
<-30 dBc at 0 dBm

4

<-45 dBc at -15 dBm

4

3rd harmonic

3

<-30 dBc at +8 dBm
<-50 dBc at 0 dBm

4

<-50 dBc at -15 dBm

4

Harmonic measurement accuracy

5

300 kHz to 3 GHz

±1.5 dB

3 GHz to 6 GHz

±3 dB (Option 006)

Harmonic measurement dynamic range

4

-40 dBc (output = -10 dBm, input = <-15 dBm)
Frequency offset mode

6

Frequency range

300 kHz to 3 GHz

Option 006

300 kHz to 6 GHz

R-channel input requirements

Power level
300 kHz to 3 GHz

0 to -35 dBm

3 GHz to 6 GHz

0 to -30 dBm

LO Spectral purity

Maximum spurious input <-25 dBc
Residual FM

<20 kHz

LO frequency accuracy

6

±1 MHz at nominal frequency

Agilent 8753ET Specifications

1. Relative to -5 dBm output power for the Agilent 8753ET; -10 dBm output power

for 8753ET with Option 004.

2. 16 MHz to 3 GHz.
3. 16 MHz to 2 GHz.
4. Typical performance.
5. 25° C ± 5° C.
6. The Agilent 8753ET source characteristics and measurement accuracy in this

mode are dependent on the stability of the external LO source. The RF source
tracks the LO to maintain a stable IF signal at the R-channel receiver input.
Degradation in accuracy is negligible when using an Agilent 8642A/B, 8656B,
or E4432B RF signal generator as the LO source.

Advertising