Measurement port characteristics – Agilent Technologies 8753ET User Manual

Page 5

Advertising
background image

5

Agilent 8753ES (50-ohm systems)

Type-N test ports

Configuration
Network analyzer

8753ES
Standard and Option 006

Calibration kit

85032B

Test-port cables

11857D

System dynamic range
30 kHz to 50 kHz

70 dB

1

50 kHz to 300 kHz

90 dB

1

300 kHz to 16 MHz

100 dB

16 MHz to 3 GHz

110 dB

Option 014

108 dB

3 GHz to 6 GHz

105 dB

Option 014

103 dB

Measurement Port Characteristics

Frequency Range

Corrected

30 kHz–

300 kHz–

1.3 GHz–

3 GHz–

300 kHz

2

1.3 GHz

3 GHz

6 GHz

Directivity

50 dB

50 dB

47 dB

40 dB

Source match 49 dB

42 dB

36 dB

31 dB

Load match

50 dB

50 dB

47 dB

40 dB

Reflection

±(0.005 dB

±(0.009 dB

±(0.019 dB

±(0.070 dB

tracking

+0.02 dB/°C) +0.01 dB/°C) +0.02 dB/°C) +0.03 dB/°C)

Transmission ±(0.014 dB

±(0.013 dB

±(0.026 dB

±(0.065 dB

tracking

+0.02 dB/°C) +0.01 dB/°C) +0.02 dB/°C) +0.03 dB/°C)

Transmission uncertainty

1. Typical below 300 kHz.
2. Typical performance.

Magnitude

Phase

Magnitude

Phase

Reflection uncertainty

Advertising
This manual is related to the following products: