System performance summaries, Measurement port characteristics – Agilent Technologies 8753ET User Manual

Page 6

Advertising
background image

6

Agilent 8753ES (50-ohm systems)

3.5-mm test ports

Configuration
Network analyzer

8753ES
Standard and Option 006

Calibration kit

85033D

Test-port cables

11857D

System dynamic range
30 kHz to 50 kHz

70 dB

1

50 kHz to 300 kHz

90 dB

1

300 kHz to 16 MHz

100 dB

16 MHz to 3 GHz

110 dB

Option 014

108 dB

3 GHz to 6 GHz

105 dB

Option 014

103 dB

Measurement Port Characteristics

Frequency Range

Corrected

30 kHz–

300 kHz–

1.3 GHz–

3 GHz–

300 kHz

2

1.3 GHz

3 GHz

6 GHz

Directivity

49 dB

46 dB

44 dB

38 dB

Source match 49 dB

44 dB

41 dB

37 dB

Load match

49 dB

46 dB

44 dB

38 dB

Reflection

±(0.010 dB

±(0.005 dB

±(0.007 dB

±(0.009 dB

tracking

+0.02/°C)

+0.01/°C)

+0.02/°C)

+0.03/°C)

Transmission ±(0.016 dB

±(0.014 dB

±(0.022 dB

±(0.048 dB

tracking

+0.02/°C)

+0.01/°C)

+0.02/°C)

+0.03/°C)

System Performance Summaries

1. Typical below 300 kHz.
2. Typical performance.

Transmission uncertainty

Magnitude

Phase

Magnitude

Phase

Reflection uncertainty

Advertising