System performance summaries, Measurement port characteristics – Agilent Technologies 8753ET User Manual

Page 8

Advertising
background image

8

Agilent 8753ES (75-ohm systems)

Type-F test ports

Configuration
Network analyzer

8753E Option 075

Calibration kit

85039B

Test-port cables

11857B

System dynamic range
30 kHz to 50 kHz

68 dB

1

50 kHz to 300 kHz

90 dB

1

300 kHz to 16 MHz

96 dB

16 MHz to 3 GHz

106 dB

Measurement Port Characteristics

Data is shown for a Type-F female reflection port and
a Type-F male transmission port.

Frequency Range

Corrected

2

30 kHz–300 kHz

300 kHz–1.3 GHz 1.3 GHz–3 GHz

Directivity

45 dB

45 dB

40 dB

Source match

40 dB

40 dB

30 dB

Load match

45 dB

45 dB

40 dB

Reflection

±(0.060 dB

±(0.060 dB

±(0.024 dB

tracking

+0.02 dB/°C)

+0.01 dB/°C)

+0.02 dB/°C)

Transmission

±(0.033 dB

±(0.019 dB

±(0.057 dB

tracking

+0.02 dB/°C)

+0.01 dB/°C)

+0.02 dB/°C)

System Performance Summaries

1. Typical below 300 kHz.
2. Typical performance.

Transmission uncertainty

Magnitude

Phase

Magnitude

Phase

Reflection uncertainty