General Conditions

RFI and EMI susceptibility: defined by VDE 0730,
CISPR Publication 11, and FCC Class B Standards.

ESD (electrostatic discharge): must be eliminated
by use of static-safe work procedures and an anti-
static bench mat. The flexible rubber keypad pro-
tects key contacts from dust, but the environment
should be as dust-free as possible for optimal relia-
bility.

Operating conditions
Temperature

0° to 55° C

(unless otherwise noted)
Humidity

5% to 95% at 40° C
(non-condensing)

Altitude

0 to 4500 meters
(15,000 feet)

Non-operating storage conditions
Temperature

-40° C to +70° C

Humidity

0 to 90% relative at +65° C
(non-condensing)

Altitude

0 to 15,240 meters
(50,000 feet)

Weight (8753ET/ES)
Net

21 kg (46 lb)

Shipping

35 kg (77 lb)

Cabinet Dimensions (8753ET/ES)

(These dimensions exclude front and rear panel
protrusions.)

222 mm H x 425 mm W x 457 mm D
(8.75 in x 16.75 in x 18.0 in)

The following table shows typical measurement
times in milliseconds.

Typical Time for Completion (msec)

Number of Points

Measurement

51

201

401

1601

Start=1 GHz, Span=10 MHz, IFBW=6 kHz

Uncorrected, 1-port calibration

1

32

70

121

423

Two-port calibration

2

62

139

240

848

Start=30 kHz, Stop=3 GHz, IFBW=6 kHz

Uncorrected, 1-port calibration

1

202

270

304

615

Two-port calibration

2

402

540

607

1237

Start=30 kHz, Stop=6 GHz, IFBW=6 kHz

Uncorrected, 1-port calibration

1

310

380

415

658

Two-port calibration

2

618

757

829

1315

Time domain
(increase over uncorrected sweep time)

3

Conversion

12

42

86

378

Gating

14

40

80

349

GPIB data transfer

4

Internal binary

10

16

21

58

ASCII

35

112

214

831

IEEE 754 floating point format

32-bit

11

19

28

83

64-bit

13

26

42

141

Options

Harmonic measurements (Option 002)
Measures amplifier 2nd and 3rd harmonics on
a swept-frequency basis for fundamental signals
above 16 MHz.
Dynamic Range (source at +10 dBm, receiver <+30 dBm)

+40 dBc (minimum)

Accuracy

5

±1 dB (<6 GHz)

Step attenuator (Option 004)
Provides source output power range from
-85 to +10 dBm for 8753ET model.

6 GHz operation (Option 006)
With Option 006, performance is extended to
6 GHz. When external source, tuned receiver or
harmonic mode is used, the receiver is capable
of measuring signals up to 6 GHz.

Time domain (Option 010)
Transforms data from the frequency domain to the
time domain using a Fourier transformation technique.

High-stability frequency reference (Option 1D5)
This option adds an ovenized 10-MHz frequency
reference output. It is connected to the external
reference input on the rear panel. See the “General
Characteristics” section for below specifications.

Environmental Characteristics

Measurement Throughput Summary

1. One-port calibration, with a 6 kHz IF bandwidth. Includes system retrace time,

but does not include bandswitch time. Time-domain gating is assumed off.

2. Same as footnote 1, but for an S21 measurement with full two-port calibration.

Includes RF switching time.

3. Option 010 only, gating off.
4. Measured with an Omnibook 7100 266 MHz Pentium

®

II computer.

5. Does not include error from the 8753T/ES source and receiver harmonics.

Agilent 8753ET/ES
Physical dimensions

Advertising
This manual is related to the following products: